Adjust the settings for nose-tail base detection (Contour-based)

Aim

To achieve detection of the nose- and tail-base points of your subjects with high accuracy.

Prerequisites

In the Experiment Settings, under Tracked Features, you chose Center-point, nose-point and tail-base detection.

You have optimized the lighting and the background. See Optimize nose-tail base detection

Procedure

1.Select the image source and sample rate. See Video file, image quality and sample rate.

2.Use the Automated Setup. When the Automated Setup gives good detection, proceed with acquisition. If not, follow the steps below. See Detection settings: Automated setup

3.Under Video, choose the sample rate.

Use a sample rate as high as possible. However, for nose-tail tracking in combination with marker-assisted tracking, use a sample rate of 12.5, 15 or 14.98 samples/s depending on the camera.

4.Open the Advanced detection settings: Method. Choose the detection method and the method for estimating the body points. See Methods for estimating the body points

If you followed the Automated Setup, a Default method is selected automatically for one-subject tracking, and For occlusions for multiple-subject tracking. See Methods for estimating the body points

In the picture below: Left, a sub-optimal result of body detection (part of the right side of the body is not detected). Right, the result when the contrast settings are optimized; now the complete body is detected. The color of the body contour at this stage is orange (=noise) because the model parameters have not been configured yet.

inset_5500336.jpg 

Notes

The maximum sample rate depends on the camera.